High-speed microplate spectrophotometer with UV-Vis wavelength range 200-999 nm, 4-Zone™ temperature control, and dual plate format support

  • Wavelength range: 200 nm - 999 nm without parasitic filter requirement
  • Ultra-fast reading: 96-well plates in 5 seconds, 384-well plates in 11 seconds
  • Advanced 4-Zone™ incubation and temperature control (37°C to 50°C ±0.2°C)
  • Compact, automation-compatible footprint (21.6 x 40.6 x 21.6 cm)
  • Integrated Gen5™ Microplate Reader Software for data collection and advanced analysis

Kargolama ve Teslimat

  • Kurye Teslimatı

Kuryemiz belirtilen adrese teslimat yapacaktır.

2-3 Gün

Ü.A. Kargo

  • Aras Kargo Teslimatı

Aras Kargo kuryesi belirtilen adrese teslimat yapacaktır.

1-3 Gün

Ücretsiz Kargo

  • Garanti 2 yıl
  • 14 Günlük ücretsiz iade

The BioTek PowerWave HT Microplate Spectrophotometer delivers exceptional plate reading speeds and a comprehensive UV-Vis wavelength range, making it ideal for diverse applications from direct quantification of nucleic acids to fully integrated high-throughput screening operations requiring speed, accuracy, and reliability. Engineered with robust hardware, proven optical performance, and a compact footprint, the PowerWave HT provides maximum flexibility while optimizing bench space in both standalone and automated system configurations.

Technical Specifications

Detection Mode Absorbance
Reading Methods Endpoint, kinetic, spectral scan
Microplate Format Support 96-well and 384-well plates
Temperature Control 4-Zone™ incubation: 37°C to 50°C ±0.2°C
Agitation Mode Linear shaking
Software Gen5™ Data Analysis Software; Gen5 21 CFR Part 11 compliant option available
Automation Compatibility BioStack and third-party automation compatible
Light Source Xenon flash
Detector Photodiode
Wavelength Selection Monochromator
Wavelength Range 200 to 999 nm in 1 nm increments
Monochromator Bandwidth 5 nm
Dynamic Range 0 – 4.0 OD
Resolution 0.001 OD
Path Length Correction Yes
Wavelength Accuracy ±2 nm
Wavelength Repeatability ±0.2 nm
OD Accuracy ±1% ±0.01 OD
OD Linearity ±1%
OD Repeatability ±0.5% ±0.005 OD
Stray Light 0.03% at 230 nm
Reading Speed (Kinetic) 96-well: 5 seconds; 384-well: 11 seconds
Power Requirements 100-240 Volt AC, 50/60 Hz
Dimensions 8.5" W x 16" D x 8.5" H (21.6 x 40.6 x 21.6 cm)
Weight 24 lbs (11 kg)
Regulatory Compliance CE and TÜV marked; RoHS compliant; In Vitro Diagnostic use models available

Applications

  • Cellular and microbial growth kinetics
  • ELISA assays and immunoassay applications
  • Nucleic acid quantification (DNA/RNA)
  • Fast kinetic and UV-Vis measurements
  • High-throughput screening in automated systems
  • Protein concentration determination
  • Enzymatic activity measurements

Frequently Asked Questions

What wavelength range does the PowerWave HT cover?
The PowerWave HT provides comprehensive UV-Vis coverage from 200 nm to 999 nm in 1 nm increments, without requiring parasitic filters.
How fast can the PowerWave HT read microplates?
Reading speeds are exceptional: 96-well plates in 5 seconds and 384-well plates in 11 seconds in kinetic mode.
What is the temperature control capability?
The advanced 4-Zone™ temperature control system maintains incubation temperatures from 37°C to 50°C with ±0.2°C precision, with integrated linear shaking capability.
Is the PowerWave HT suitable for automation?
Yes, the compact footprint (21.6 x 40.6 x 21.6 cm) and compatibility with BioStack and third-party automation systems make it ideal for integrated high-throughput workflows.
What software is provided?
The Gen5™ Microplate Reader Software is included for data collection, analysis, and reporting. A 21 CFR Part 11 compliant version is available upon request.
What is the optical accuracy and repeatability?
OD accuracy is ±1% ±0.01 OD with ±0.5% ±0.005 OD repeatability. Wavelength accuracy is ±2 nm with ±0.2 nm repeatability.