Stoklu siparişleriniz aynı gün kargoda!

Climate Chamber Binder KBF 240 (247L)

Fiyat Sorunuz

Precision climate chamber with temperature control 0 °C to 70 °C, humidity range 10% to 80% RH, 247 L capacity, APT.line™ pre-heating chamber technology, and comprehensive safety features.

  • Temperature range 0 °C to 70 °C with ±0.1 K fluctuation
  • Humidity range 10% to 80% RH with ≤1.5% RH precision
  • APT.line™ pre-heating chamber technology for rapid stabilization
  • Capacitive humidity sensor with vapor humidification system
  • Intuitive touchscreen controller with time-segment and real-time programming
  • Integrated data logger with USB readout in open format
  • Hermetic safety glass (ESG) interior door with TIMELESS anti-corrosion coating
  • Stainless steel interior chamber with 2 stainless steel shelves
  • Class 3.1 independent temperature safety device with visual and audible alarm (DIN 12880)
  • Silicon plug access port (30 mm) on left side
  • Door heating for condensation prevention
  • Ethernet interface for computer integration
  • Self-diagnostic unit health analysis
  • Mobility with 4 fixed wheels (2 braked) for 247 L capacity
  • Nominal power 2.1 kW, single-phase 200–240 V

Kargolama ve Teslimat

  • Kurye Teslimatı

Kuryemiz belirtilen adrese teslimat yapacaktır.

2-3 Gün

Ü.A. Kargo

  • Aras Kargo Teslimatı

Aras Kargo kuryesi belirtilen adrese teslimat yapacaktır.

1-3 Gün

Ücretsiz Kargo

  • Garanti 2 yıl
  • 14 Günlük ücretsiz iade

The Binder KBF 240 is a versatile climate chamber engineered for demanding laboratory and industrial applications requiring precise environmental simulation. Featuring an expansive temperature range of 0 °C to 70 °C and relative humidity control from 10% to 80% RH, this unit delivers exceptional stability with temperature fluctuations of ±0.1 K and humidity precision of ≤1.5% RH. The APT.line™ pre-heating chamber technology ensures rapid thermal stabilization, while capacitive humidity sensors combined with vapor humidification enable rapid humidity recovery. The intuitive touchscreen controller supports both time-segment and real-time programming, complemented by an integrated data logger with USB readout. Built with a hermetic safety glass interior door featuring TIMELESS anti-corrosion coating and a stainless steel chamber, the unit includes 2 stainless steel shelves, Class 3.1 independent temperature safety device (DIN 12880), door heating, and Ethernet connectivity for seamless laboratory integration.

Technical Specifications

Order Number 9020-0323
Temperature Range 0 … 70 °C
Temperature Fluctuation [± K] 0.1
Temperature Change at 40 °C [± K] 0.3
Maximum Heat Compensation at 40 °C [W] 300
Humidity Range 10 … 80 % RH
Humidity Fluctuation at 25 °C and 60% RH ≤1.5 ±% RH
Humidity Fluctuation at 40 °C and 75% RH ≤1.5 ±% RH
Humid Warm-up Time at 25 °C and 60% RH after 30 s door opening [min] 5
Humidity Recovery Time at 40 °C and 75% RH after 30 s [min] 11
Nominal Power [kW] 2.1
Phase (Nominal Voltage) 1 ~
Power Frequency [Hz] 50/60
Rated Voltage [V] 200 … 240
Unit Fuse [A] 16
Interior Doors 1
Unit Doors 1
Interior Width [mm] 650
Interior Height [mm] 785
Interior Depth [mm] 485
Interior Volume [L] 247
Load Capacity per Shelf [kg] 30
Number of Stainless Steel Shelves 2
Access Port Silicon plug, 30 mm, left side
Wheels 4 fixed wheels (2 braked)
Computer Interface Ethernet
Temperature Safety Device Class 3.1 independent, visual and audible alarm (DIN 12880)

Applications

  • Pharmaceutical stability testing and storage
  • Biological sample preservation and microbial culture incubation
  • Material science research and accelerated aging studies
  • Electronic component testing and temperature cycling
  • Environmental chamber for product qualification and validation
  • Food preservation testing and shelf-life studies
  • Polymer and adhesive degradation analysis
  • Climate simulation for reliability engineering

Frequently Asked Questions

What is the temperature stability of the KBF 240?
The unit achieves a temperature fluctuation of ±0.1 K, ensuring excellent stability for sensitive experiments. Temperature change at 40 °C is maintained at ±0.3 K, meeting stringent laboratory standards.
How quickly does the chamber stabilize humidity after door opening?
At 25 °C and 60% RH, humidity recovers within 5 minutes after 30 seconds of door opening. At 40 °C and 75% RH, recovery takes approximately 11 minutes, thanks to the vapor humidification system.
What data logging capabilities are included?
The integrated data logger records measured values that can be read via USB in open format, facilitating data export and compliance documentation for regulated environments.
Is the KBF 240 suitable for multi-stage temperature protocols?
Yes. The touchscreen controller supports both time-segment and real-time programming, enabling complex multi-phase temperature and humidity protocols for research and validation studies.
What safety features protect against temperature excursions?
The unit includes a Class 3.1 independent temperature safety device (DIN 12880) with visual and audible alarms, ensuring immediate notification of temperature deviations. Door heating prevents condensation on the inner glass.
Can the KBF 240 be integrated into laboratory information systems?
Yes. The Ethernet interface enables computer connectivity and integration with laboratory monitoring and data management systems.
What is the interior chamber material?
The interior is constructed from stainless steel with a hermetic safety glass (ESG) door featuring TIMELESS anti-corrosion coating to prevent long-term glass degradation in humid environments.
How many shelves are included and what is the load capacity?
The unit includes 2 stainless steel shelves, each rated for a maximum load of 30 kg, providing flexible sample organization within the 247 L working volume.