Stoklu siparişleriniz aynı gün kargoda!

Atago CM-BASEα-88S In-line Propylene Glycol Monitor (In-line Refractometer)

Fiyat Sorunuz

  • Concentration measurement range: 0.0-90.0%, accuracy ±0.4%
  • Freeze point measurement: -50°C to 0°C, accuracy ±1°C
  • Wide ATC range: 10-95°C, temperature precision ±0.2°C
  • 4-20mA and RS-232C output options for seamless system integration
  • Second popular model in the RRI-approved CM-BASEα series
  • No IP65 protection — enclosed/controlled environment installation required

Kargolama ve Teslimat

  • Kurye Teslimatı

Kuryemiz belirtilen adrese teslimat yapacaktır.

2-3 Gün

Ü.A. Kargo

  • Aras Kargo Teslimatı

Aras Kargo kuryesi belirtilen adrese teslimat yapacaktır.

1-3 Gün

Ücretsiz Kargo

  • Garanti 2 yıl
  • 14 Günlük ücretsiz iade

CM-BASEα-88S is an Atago brand refractometric monitor designed for continuous in-line measurement of propylene glycol concentration and freeze point in HVAC cooling systems, food/beverage processing, and industrial process lines. The instrument delivers real-time data to operators through direct measurement without sampling from the process. The unit measures concentration in the 0.0-90.0% range with ±0.4% accuracy and freeze point between -50°C and 0°C with ±1°C accuracy; both values are calculated from a single sensor. A wide 10-95°C ATC (automatic temperature compensation) range with ±0.2°C precision ensures stable measurement across variable process temperatures. 4-20mA and RS-232C outputs enable seamless integration with SCADA/PLC systems. The product lacks IP65 protection, making it suitable for controlled, enclosed environment installations; additional enclosure may be required in open/humid fields. For HVAC chiller and heat pump systems requiring freeze protection, food-beverage industries needing food-grade PG control, and chemical/process cooling lines requiring continuous monitoring, this instrument offers a practical solution. As the second popular model in the RRI-approved CM-BASEα series, it adapts proven refractometric measurement technology for in-line application.

Technical Specifications

Manufacturer Atago
Model / Catalog No. CM-BASEα-88S / 3778
Concentration Range 0.0-90.0%
Concentration Resolution 0.1%
Concentration Accuracy ±0.4%
Freeze Point Range -50~0°C
Freeze Point Resolution 1°C
Freeze Point Accuracy ±1°C
ATC (Temperature Compensation) Range 10-95°C
ATC Precision ±0.2°C
Output 4-20mA, RS-232C
IP Protection Class No IP65

Applications

  • Freeze protection control in HVAC chiller and heat pump systems
  • Food-grade propylene glycol cooling lines in food and beverage industries
  • Continuous concentration/freeze point monitoring in industrial process cooling systems

Frequently Asked Questions

Does the CM-BASEα-88S measure both concentration and freeze point simultaneously?
Yes, refractive measurement acquired from a single in-line sensor simultaneously calculates both propylene glycol concentration (%) and freeze point (°C).
What does the absence of IP65 protection mean?
The instrument does not provide IP65-level protection against water splash or dust; additional enclosure or installation within a closed control panel is recommended for open field or humid/dusty environments.
What is the practical benefit of the 10-95°C ATC range?
It provides automatic compensation with ±0.2°C precision across a broad temperature span from cold storage to high-temperature process lines, enabling a single unit to cover diverse process conditions.