Stoklu siparişleriniz aynı gün kargoda!

Biotek Epoch Microplate Spectrophotometer (ELISA Reader)

Fiyat Sorunuz

Monochromator-based, filter-free absorbance microplate reader. Cost-effective solution for ELISA, nucleic acid, and protein measurements in the 200-999 nm wavelength range. Micro-volume measurement (2 µL) possible with Take3 plate.

  • 200-999 nm monochromator-based filter-free wavelength selection (1 nm increments)
  • 0-4.0 OD dynamic range, 0.0001 OD resolution
  • Xenon flash light source and photodiode detector
  • 6-384 well plate support, 2 µL micro-volume measurement with Take3
  • Endpoint, kinetic, wavelength scan, and well area scan reading modes
  • Controlled by Gen5 Software, BioStack and third-party automation compatible

Kargolama ve Teslimat

  • Kurye Teslimatı

Kuryemiz belirtilen adrese teslimat yapacaktır.

2-3 Gün

Ü.A. Kargo

  • Aras Kargo Teslimatı

Aras Kargo kuryesi belirtilen adrese teslimat yapacaktır.

1-3 Gün

Ücretsiz Kargo

  • Garanti 2 yıl
  • 14 Günlük ücretsiz iade

Biotek Epoch is a microplate spectrophotometer that performs absorbance measurements across a wide wavelength range without requiring filters, thanks to its monochromator-based optical design. It is a practical and low-maintenance system for laboratories conducting ELISA analyses, nucleic acid quantification, and protein measurements.

The instrument supports multiple reading modes including endpoint, kinetic, wavelength scan, and well area scan; compatible with 6 to 384-well microplates. With the optional Take3 Micro-Volume Plate, UV-Vis measurements are possible with sample volumes as low as 2 µL per well. The combination of xenon flash light source and photodiode detector enables wavelength selection in 1 nm increments across the 200-999 nm range.

Controlled by Gen5 Data Analysis Software, Epoch offers simple programming and powerful data analysis capabilities; optional Gen5 Secure provides 21 CFR Part 11 compliance. An 'R' configuration compatible with BioStack and third-party automation systems is available. This model, which lacks temperature control and shaking features, is positioned as a cost-effective alternative for fundamental absorbance measurement requirements.

Technical Specifications

Detection mode Absorbance
Reading modes Endpoint, kinetic, wavelength scan, well area scan
Microplate types 6-384 well plates
Supported accessories Take3 Micro-Volume Plate
Temperature control Not available
Shaking Not available
Software Gen5 Data Analysis Software (optional Gen5 Secure – 21 CFR Part 11)
Automation compatibility BioStack and third-party automation ('R' configuration)
Light source Xenon flash
Detector Photodiode
Wavelength selection Monochromator
Wavelength range 200-999 nm (1 nm increments)
Monochromator bandwidth 5 nm
Dynamic range 0-4.0 OD
Resolution 0.0001 OD
Optical path length correction Available
Monochromator wavelength accuracy ±2 nm
Monochromator wavelength repeatability ±0.2 nm
OD accuracy 0-2 OD: ±1% +0.010 OD; 2-2.5 OD: ±3% +0.010 OD

Applications

  • ELISA absorbance measurements
  • Nucleic acid quantification (micro-volume with Take3)
  • Protein concentration measurement
  • Kinetic and endpoint absorbance assays
  • Spectral analysis requiring wavelength scanning

Frequently Asked Questions

What is the minimum sample volume measurable with Epoch?
With the optional Take3 Micro-Volume Plate, UV-Vis absorbance measurements can be performed with sample volumes as low as 2 µL per well.
Does the Epoch have temperature control or shaking capabilities?
No, the Epoch model does not include temperature control or shaking (orbital/linear motion) features.
What software controls the Epoch?
The instrument is controlled by Gen5 Data Analysis Software; optional Gen5 Secure can provide 21 CFR Part 11 compliance.