Stoklu siparişleriniz aynı gün kargoda!

Binder KBF 1020 Climate Conditioning Chamber (1020L)

Fiyat Sorunuz

Laboratory climate chamber with temperature control 0–70°C and humidity range 10–80% RH, featuring APT.line™ pre-heating technology and intuitive touch screen controller.

  • Temperature range: 0°C – 70°C with ±0.5 K fluctuation
  • Humidity range: 10% – 80% RH with ≤1.5% RH stability
  • APT.line™ pre-heating chamber technology for rapid stabilization
  • Capacitive humidity sensors with steam humidification system
  • Intuitive touch screen controller with time-segment and real-time programming
  • Integrated data logger; measured values readable via USB in open format
  • Hermetic interior door constructed from ESG safety glass
  • TIMELESS coating prevents glass corrosion
  • Stainless steel interior chamber with two stainless steel shelves
  • Silicone-plugged access port, 30 mm, left side
  • Four fixed wheels with two locking wheels (capacity from 240 liters)
  • Class 3.1 independent temperature safety device (DIN 12880) with visual and audible alarm
  • Ethernet computer interface
  • Door heating element
  • Unit self-test for comprehensive status analysis

Kargolama ve Teslimat

  • Kurye Teslimatı

Kuryemiz belirtilen adrese teslimat yapacaktır.

2-3 Gün

Ü.A. Kargo

  • Aras Kargo Teslimatı

Aras Kargo kuryesi belirtilen adrese teslimat yapacaktır.

1-3 Gün

Ücretsiz Kargo

  • Garanti 2 yıl
  • 14 Günlük ücretsiz iade

The Binder KBF 1020 is a precision climate conditioning chamber designed for advanced material testing, sample storage, and climate simulation in research laboratories. With a 1020-liter capacity, this unit delivers reliable temperature and humidity control across a wide operating range. The APT.line™ pre-heating chamber technology ensures rapid stabilization and uniform climate conditions throughout the chamber. The capacitive humidity sensors combined with steam humidification provide precise humidity management. An intuitive touch screen controller enables time-segment and real-time programming, while the integrated data logger records all measured values readable via USB in open format. The stainless steel interior construction with ESG safety glass door and TIMELESS coating prevents glass corrosion. The unit includes self-diagnostic capabilities for comprehensive system analysis and is equipped with a Class 3.1 independent temperature safety device (DIN 12880) with visual and audible alarms.

Technical Specifications

Order number 9020-0326
Max. heat compensation at 40°C 600 W
Temperature fluctuation ±0.5 K
Temperature range 0 – 70°C
Temperature deviation at 40°C ±0.2 K
Humidity fluctuation at 25°C and 60% RH ≤1.5 ±% RH
Humidity fluctuation at 40°C and 75% RH ≤1.5 ±% RH
Humidity range 10 – 80% RH
Humidity rise time (30 s after door opening) at 25°C and 60% RH 11 min
Humidity fall time (30 s after door opening) at 40°C and 75% RH 12 min
Temperature fluctuation at 25°C and 60% RH ±0.1 K
Temperature fluctuation at 40°C and 75% RH ±0.1 K
Temperature deviation at 25°C and 60% RH ±0.2 K
Temperature deviation at 40°C and 75% RH ±0.2 K
Rated power 3.1 kW
Power supply phase 1 ~
Frequency 50/60 Hz
Nominal voltage 200 – 230 V
Unit fuse 16 A
Interior doors 2
Unit doors 2
Interior depth 836 mm
Interior height 1250 mm
Interior width 976 mm
Internal volume 1020 L
Load per shelf 45 kg

Applications

  • Material testing under controlled climate conditions
  • Stability testing of pharmaceutical products and formulations
  • Accelerated aging studies for polymers and composites
  • Environmental simulation for electronic component testing
  • Biological sample storage and incubation
  • Quality control testing in chemical laboratories
  • Research applications requiring precise temperature and humidity control
  • Climate chamber simulation for product development

Frequently Asked Questions

What is the APT.line™ pre-heating technology?
APT.line™ is Binder's proprietary pre-heating chamber technology that enables rapid stabilization of temperature and humidity conditions, reducing conditioning time and improving test efficiency.
Can the humidity be controlled independently of temperature?
Yes. The chamber features capacitive humidity sensors combined with steam humidification, allowing independent humidity regulation across the full operating range of 10–80% RH.
How are measured values recorded and retrieved?
The integrated data logger automatically records all measured values, which can be read in open format via USB connection for easy data analysis and archiving.
What safety features does this chamber include?
The unit is equipped with a Class 3.1 independent temperature safety device (DIN 12880) featuring visual and audible alarms, ESG safety glass door, and comprehensive self-diagnostic capabilities.
Is the chamber mobile?
Yes, the unit is equipped with four fixed wheels, two of which have locking capability, allowing for convenient repositioning while maintaining stability during operation.
What interface options are available for data transfer?
The chamber features an Ethernet computer interface for network connectivity and integrated USB access for direct data logger retrieval in open format.
What is the maximum load per shelf?
Each of the two stainless steel shelves supports a maximum load of 45 kg, for a total shelf capacity of 90 kg.
Does the chamber include temperature stabilization features?
Yes. The APT.line™ pre-heating technology combined with the 600 W heating capacity at 40°C and door heating element ensures rapid and uniform temperature stabilization throughout the chamber.