Stoklu siparişleriniz aynı gün kargoda!

Refrigerated Incubator Binder KB400 Series 400L

Fiyat Sorunuz

High-precision cooled incubator with electronic temperature control, -5 to 100°C range, 400L capacity

  • Guaranteed temperature accuracy and reproducible results with electronic APT.line™ controlled heating chamber
  • Temperature range from 5°C (32°F) ambient to 100°C (212°F) with exceptional stability
  • Integrated timer 0-99 hours with DS controller for automated operation
  • Digital temperature control with 1°C precision adjustment
  • Independent adjustable over-temperature safety device with visual alarm, Class 2 (DIN 12880)
  • Adjustable pre-ventilation front cover and rear exhaust ø 50 mm (1.97 in.)
  • Stackable up to 115L units for flexible space optimization
  • Includes 2 chrome-plated shelves for sample organization
  • BINDER factory test certification for quality assurance

Kargolama ve Teslimat

  • Kurye Teslimatı

Kuryemiz belirtilen adrese teslimat yapacaktır.

2-3 Gün

Ü.A. Kargo

  • Aras Kargo Teslimatı

Aras Kargo kuryesi belirtilen adrese teslimat yapacaktır.

1-3 Gün

Ücretsiz Kargo

  • Garanti 2 yıl
  • 14 Günlük ücretsiz iade

The Binder KB400 Series 400L refrigerated incubator delivers precise temperature control and reproducible results for cell culture, microbiology, and quality control applications. Featuring electronically controlled APT.line™ heating chamber with ±0.1K stability at 37°C, integrated timer (0-99 hours), and independent over-temperature safety device. Stackable design with integrated chrome-plated shelving and BINDER test certification ensures reliable performance in demanding laboratory environments.

Technical Specifications

Model Option Standard
Order Number 9020-0203
Recovery time after door opening 30 sec at 37°C 4 min
Temperature fluctuation at 37°C ±0.1 K
Temperature range -5 to 100°C
Temperature deviation at 37°C ±0.2 K
Nominal power 1.4 kW
Phase (Nominal voltage) 1 ~
Power frequency 50/60 Hz
Rated voltage 200 … 240 V
Unit fuse 16 A
Inner doors 1
Unit doors 1
Internal depth 485 mm
Internal height 1270 mm
Internal width 650 mm
Internal volume 400 L
Load per shelf 30 kg
Unit net weight (empty) 221 kg
Maximum permitted load 120 kg
External depth 805 mm
External height 1950 mm
Wall clearance rear 100 mm
Wall clearance side 100 mm
External width 925 mm
Energy consumption at 37°C 330 Wh/h
Sound pressure level 53 dB(A)
Number of shelves (standard/max.) 2/2

Applications

  • Cell and tissue culture incubation
  • Microbiology and bacterial culture applications
  • Pharmaceutical and quality control testing
  • Biochemical assay and enzyme research
  • Temperature-dependent sample storage and processing
  • Stability testing and shelf-life studies

Frequently Asked Questions

What is the minimum operating ambient temperature?
The KB400 Series can operate in ambient temperatures as low as 5°C (32°F). For optimal performance and faster recovery times, ambient conditions should be maintained between 16-27°C.
How quickly does the chamber recover temperature after door opening?
The recovery time after 30 seconds of door opening at 37°C is approximately 4 minutes, ensuring minimal disruption to ongoing incubations.
Can multiple KB400 units be stacked?
Yes, KB400 units can be stacked vertically up to a combined capacity of 115L (utilizing the stackable frame system). This allows for flexible space optimization in laboratory environments.
What safety features are included?
The unit features an independent, adjustable over-temperature safety device (Class 2, DIN 12880) with visual alarm indication, preventing overheating scenarios and protecting samples.
Is the KB400 certified?
Yes, each Binder KB400 Series incubator includes factory test certification (BINDER test certificate) confirming compliance with specifications and quality standards.
What are the electrical requirements?
The KB400-230V requires single-phase 230V (200-240V rated), 50/60 Hz operation with a 16A fuse. Nominal power consumption is 1.4 kW.