Stoklu siparişleriniz aynı gün kargoda!

Laboratory Vacuum Chamber 53L with Integrated Diaphragm Pump System

Fiyat Sorunuz

Precision temperature-controlled vacuum chamber (53L) with real-time monitoring, programmable temperature profiles, and VACUUBRAND MZ 2C NT chemistry diaphragm pump for materials testing and sample conditioning.

  • Temperature range: ambient +15 to 200°C with ±1.5 K stability at 100°C
  • Chamber volume: 53 liters with aluminum shelving system (2 standard, expandable to 4)
  • Integrated VACUUBRAND MZ 2C NT two-stage chemistry diaphragm pump
  • Minimum vacuum specification: 0.01 mbar with 0.01 bar/hour leak rate
  • Real-time temperature monitoring and time-step programmable control
  • RS 422 computer interface for automated process control and data logging
  • Inert gas purge capability with flow restrictor (3/8" connection)
  • Rapid temperature ramp: 65 minutes to reach 100°C
  • Class 2 protection classification with visual alarm system
  • Compact internal dimensions: 285 × 285 × 295 mm

Kargolama ve Teslimat

  • Kurye Teslimatı

Kuryemiz belirtilen adrese teslimat yapacaktır.

2-3 Gün

Ü.A. Kargo

  • Aras Kargo Teslimatı

Aras Kargo kuryesi belirtilen adrese teslimat yapacaktır.

1-3 Gün

Ücretsiz Kargo

  • Garanti 2 yıl
  • 14 Günlük ücretsiz iade

Industrial-grade laboratory vacuum chamber designed for thermal processing, material drying, and environmental testing applications. Features precise temperature control from ambient +15°C to 200°C, achieves vacuum levels down to 0.01 mbar, and includes integrated two-stage chemistry diaphragm pump system (VACUUBRAND MZ 2C NT). Equipped with RS 422 computer interface for automated process control, time-step programming capability, and Class 2 safety classification with visual alarm notification.

Technical Specifications

Temperature Range Ambient +15 to 200°C
Chamber Volume 53 L
Temperature Stability (at 100°C) ±1.5 K
Ramp Time to 100°C 65 minutes
Minimum Achievable Vacuum 0.01 mbar
Leak Rate 0.01 bar/hour
Vacuum Pump Model VACUUBRAND MZ 2C NT (two-stage, chemistry diaphragm)
Pump Maximum Speed 2.0 m³/hour
Pump Achievable Vacuum 7 mbar
Internal Dimensions (W × H × D) 285 × 285 × 295 mm
Shelf Configuration 2 standard / 4 maximum (aluminum)
Maximum Chamber Load Capacity 35 kg
Maximum Load per Shelf 20 kg
Vacuum Port Connection (Chamber) DN 16 small flange
Vacuum Gauge Port Connection DN 16 small flange
Inert Gas Connection 3/8" with flow restrictor
Pump Inlet Connection DN 10 mm hose
Power Consumption 0.8 kW
Electrical Specifications 230 V, 50/60 Hz, single phase, 10 A
Pump Power Supply 230 V, 50-60 Hz, CEE/CN mains cable
Pump Certification NRTL C/US
Safety Classification Class 2 with visual alarm
Control Interface RS 422 computer interface
Control Features Real-time monitoring, time-step programming

Applications

  • Material drying and moisture removal in research and quality control
  • Thermal aging and accelerated life testing of polymers and composites
  • Outgassing and vacuum baking of electronic components and assemblies
  • Sample conditioning for pharmaceutical and chemical analysis
  • Low-pressure thermal processing and degassing of liquids
  • Environmental chamber simulation for materials testing
  • Vacuum impregnation of porous materials
  • Controlled atmosphere heating for specialized manufacturing processes

Frequently Asked Questions

What is the maximum sample load capacity?
The chamber supports a maximum total load of 35 kg, with 20 kg maximum per individual aluminum shelf. The system accommodates 2 standard shelves, expandable to 4 shelves maximum.
How long does it take to reach 100°C from ambient temperature?
The system requires approximately 65 minutes to reach 100°C under standard operating conditions.
What is the temperature stability during operation?
Temperature variation is ±1.5 K at 100°C, ensuring precise and repeatable thermal conditioning.
Can the system be operated with an inert gas atmosphere?
Yes, the chamber features an inert gas connection (3/8" with integrated flow restrictor) enabling controlled atmosphere operation to prevent oxidation during heating.
What type of vacuum pump is integrated?
A VACUUBRAND MZ 2C NT two-stage chemistry diaphragm pump is integrated, achieving vacuum levels down to 7 mbar with maximum pumping speed of 2.0 m³/hour.
Is the system compatible with laboratory data acquisition systems?
Yes, the RS 422 computer interface allows integration with SCADA systems and laboratory information management systems (LIMS) for automated process control and real-time data logging.
What electrical specifications are required for installation?
The system requires 230 V, 50/60 Hz, single-phase power supply at 10 A maximum. Both chamber and integrated pump operate on standard industrial mains power.
Can shelves be reconfigured during operation?
The system comes with 2 aluminum shelves installed and supports expansion to 4 shelves maximum. Shelf configuration should be completed before bringing the chamber to vacuum.
What is the pump certification status?
The integrated VACUUBRAND pump carries NRTL (Nationally Recognized Testing Laboratory) C/US certification, confirming compliance with North American and European safety standards.