Stoklu siparişleriniz aynı gün kargoda!

POLEKO CLN 115 SMART Incubator 115L

Fiyat Sorunuz

Laboratory incubator with natural air convection, 115L chamber capacity, and intelligent temperature control system

  • 115L (112L working) chamber capacity with natural air convection
  • Temperature range: ambient +5°C to +100°C with 0.1°C resolution
  • Temperature stability ±0.2°C at 37°C (DIN 2080, Class 2.0/3.1 overheat protection)
  • 4.3-inch full-color touchscreen interface with smart phone-like operation
  • Dual-door design with observation window for non-disruptive sample access
  • Microprocessor-controlled with 6-segment temperature-time profiling
  • Flexible shelving: 2 standard to 7 maximum stainless steel wire racks
  • Maximum unit load capacity: 60 kg (120 kg with reinforced version)
  • Internal dimensions: 460mm W × 540mm H × 450mm D
  • Automatic door damping control and open-door alarm system

Kargolama ve Teslimat

  • Kurye Teslimatı

Kuryemiz belirtilen adrese teslimat yapacaktır.

2-3 Gün

Ü.A. Kargo

  • Aras Kargo Teslimatı

Aras Kargo kuryesi belirtilen adrese teslimat yapacaktır.

1-3 Gün

Ücretsiz Kargo

  • Garanti 2 yıl
  • 14 Günlük ücretsiz iade

The POLEKO CLN 115 SMART is a precision laboratory incubator designed for microbiological sample incubation, thermal resistance analysis, and specialized testing applications. Featuring a 4.3-inch full-color touchscreen display, microprocessor-based controller, and temperature range from 5°C above ambient to +100°C, this incubator provides excellent temperature stability (±0.2°C at 37°C) and uniform distribution. The unit accommodates 2-7 stainless steel wire shelves with configurable loading capacity and includes dual-door access with observation window for convenient sample monitoring without environmental disruption.

Technical Specifications

Air Convection Type Natural
Chamber Capacity [L] 112
Working Capacity [L] 112
Controller External LCD graphic microprocessor
Display 4.3" full-color touchscreen
Temperature Range [°C] Ambient +5°C to +100°C
Temperature Resolution [°C] 0.1
Temperature Stability at 37°C [°C] ±0.2
Temperature Fluctuation at 37°C [°C] 0.6
Overheat Protection DIN 2.080 Class 2.0/Class 3.1 (optional)
Door Type Dual-door with observation window (optional single-door configuration)
Interior Material Stainless steel DIN 1.4301
Interior Coating (Smart) Stainless steel linen coating
Exterior Material Powder-coated sheet steel (graphite front panel, gray finish)
External Width A [mm] 660
External Height B [mm] 850
External Depth C [mm] 710
Internal Width D [mm] 460
Internal Height E [mm] 540
Internal Depth F [mm] 450
Shelves (Standard | Maximum) 2 | 7
Maximum Shelf Load Capacity [kg] 25
Maximum Shelf Load – Reinforced Version PW [kg] 50
Maximum Unit Load Capacity [kg] 60
Maximum Unit Load – Reinforced Version W [kg] 120
Unit Weight [kg] 65
Power Supply 230V 50-60Hz
Nominal Power [W] 450

Applications

  • Microbiological sample incubation for identification and detection
  • Thermal resistance analysis of samples exposed to elevated temperatures
  • Antibody and immunological assay testing
  • Bacterial culture and sensitivity testing
  • Crystallization observation and monitoring
  • Thermophilic microorganism cultivation
  • Pharmaceutical drug stability testing
  • Food industry denaturation and preservation testing
  • Environmental sample conditioning
  • Quality control and regulatory compliance testing

Frequently Asked Questions

What is the temperature stability tolerance of this incubator?
The CLN 115 SMART maintains temperature stability of ±0.2°C at 37°C with a temperature fluctuation of 0.6°C, meeting strict laboratory requirements for sensitive applications.
Can I upgrade the loading capacity of this unit?
Yes. The incubator is available in a reinforced version (W) that increases maximum unit load capacity from 60 kg to 120 kg, and reinforced shelves (PW) that support 50 kg per shelf instead of 25 kg.
How many shelves can this incubator accommodate?
The standard configuration includes 2 stainless steel wire shelves, expandable up to 7 shelves maximum, providing flexible capacity for different sample volumes.
What overheat protection standard does this incubator comply with?
The unit complies with DIN 2.080 Class 2.0 overheat protection, with Class 3.1 protection available as an optional upgrade.
Is the interior accessible without disturbing the temperature environment?
Yes. The dual-door design with observation window allows sample monitoring and access without opening the main door, minimizing temperature fluctuations inside the chamber.
What programming capabilities does the touchscreen interface offer?
The 4.3-inch full-color touchscreen provides a user-friendly interface with 6-segment temperature-time profiling capability, allowing complex temperature protocols for specialized applications.